Oncology-Recent Advances in External Beam Radiotherapy (EBRT)
I. Improved Accuracy and Precision
A. Volume Definition:
A. Inverse Planned Intensity-Modulated Radiotherapy (IMRT):
I. Improved Accuracy and Precision
A. Volume Definition:
- Past: Planning scans relied on diagnostic scanners, resulting in less precise tumor volume definition.
- Present: High-resolution CT simulators are standard. 3D tumor volume definition is now accurate even for palliative treatments. Fusion of CT with MRI and PET scans further refines this for radical treatments.
- Past: A 1cm margin was added to the Clinical Target Volume (CTV) to account for positioning errors. Megavoltage X-rays provided low-quality confirmation images.
- Present:
- Electronic Portal Imaging Devices (EPIDs): Replaced port films, offering improved image quality. Online imaging corrects misalignments before treatment, while offline review identifies recurring setup errors.
- Intra-treatment Imaging: Precise tumor location is confirmed during treatment using:
- Fiducial markers (e.g., gold seeds)
- Ultrasound
- Cone beam CT (mounted on the linear accelerator) for 3D online imaging.
A. Inverse Planned Intensity-Modulated Radiotherapy (IMRT):
- Mechanism: Computer optimization varies radiation beam intensity to meet dose constraints for healthy organs while delivering prescribed doses to the tumor (CTV and Gross Tumor Volume - GTV). This is achieved through:
- Step and shoot: Superimposing static uniform intensity segments.
- Dynamic MLC delivery: A shaped sliding window across the field.
- Tomotherapy: Multiple arcs with intensity modulated by dynamic MLCs.
- Benefits: Reduced late effects on normal tissues with equivalent cancer control compared to conventional radiotherapy, particularly in prostate and head/neck cancers. Promising results in lung and gynecological cancers suggest it may replace 3D conformal radiotherapy (CRT) for many radical treatments.
- Drawback: Significantly increases the workload of the physics team due to the extensive time (at least 2 hours per patient) required for meticulous volume definition.
- Addresses the challenge of: Treating mobile structures (e.g., lung cancers moving with respiration).
- Method: Links 4D CT planning (tumor volume defined at each respiratory phase) to treatment delivery during the expiratory phase for optimal coverage with minimal field size.
- Established for: Intracranial conditions (benign tumors, arteriovenous malformations).
- Method: High precision (1-2mm) treatment of small lesions (<1cm3) using an external 3d coordinate system and stereotactic fixation. typically uses 1-3 large fractions (12-20 gy).< />pan>
- Recent expansion: Successful treatment of small malignancies in the brain, lung, and liver. IGRT facilitates accuracy in sites unsuitable for localization frames.
- CTV: Clinical Target Volume
- GTV: Gross Tumor Volume
- EPID: Electronic Portal Imaging Device
- IMRT: Intensity-Modulated Radiotherapy
- MLC: Multileaf Collimator
- IGRT: Image-Guided Radiotherapy
- 4D: Four-Dimensional (referring to respiratory gating)